Differential Scanning Calorimeter
Technical and Characterization Specifications:- Measures melting temperature, heat of fusion, latent heat of melting, glass transition temperature, crystalline phase transition temperature and energy, precipitation energy and temperature
- Determines the amount of energy absorbed or released by a sample when it is heated or cooled, providing quantitative and qualitative data on endothermic and exothermic processes.
- Measurements are highly quantitative.
- Works on power compensation temperature null principle.
- Comprise independent dual furnaces constructed of Platinum-Iridium alloys with independent platinum resistance.
- Comprises distributed, platinum resistance thermometers.
- Auto sampler: up to 44 samples can be measured, in series unattended.
- It has ability to be customized through pyris player to meet analysis needs.
- Range: -170°C to 730°C
- Accuracy / Precision ± 0.1°C/ ±0 .01°C.
- Temperature overshoot: 100°C / min.
- Scanning rates: Heating /cooling 0.01°C to 500°C / min.
- High pressure cell: Extends the capability of the power compensation designed to elevated pressure measurements.
- Straight forward, pure linear heating ramps and isothermal scans.
- Provides clearer information about glass transition event.
- Atmosphere: Static, dynamic including nitrogen, argon, helium carbon dioxide, air, oxygen over full temperature range.